Operation
Instrument Controls Menu
Custom O
3-28
Model 49i Instruction Manual
CUSTOM O3 LEVELS:
>LEVEL 1
LEVEL 2
LEVEL 3
LEVEL 4
LEVEL 5
RANGE
Levels
The Custom O
3
A percentage of 100% results in the maximum amount of ozone
production. A percentage of 0% results in no ozone production. The range
of percentages, however, are not linear. This screen is only displayed if the
ozonator option is installed.
•
In the Main Menu, choose Instrument Controls > Custom Levels >
Custom Level 1, 2, 3, 4, or 5.
•
Use
•
Press
CUSTOM O3 LEVEL 1:
LAMP SETTING:
RANGE
Pump
The Pump screen allows the user to toggle the pump on and off.
•
In the Main Menu, choose Instrument Controls > Pump.
•
Press
90.0 %
75.0 %
50.0 %
25.0 %
10.0 %
AVG
DIAGS ALARM
Levels screen is used to change the custom ozonator levels.
3
and
to increment or decrement the numeric value.
to save the new concentration setting.
10.0% ?
INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
to toggle and set the pump on or off.