Communications Configuration - Thermo Scientific 49i Instruction Manual

Uv photometric o3 analyzer
Table of Contents

Advertisement

Communications
Configuration
Thermo Fisher Scientific
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example that follows saves the parameters to FLASH
memory.
Send:
set save params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example that
follows reports that the internal time is 2:15:30 pm.
Send:
time
Receive:
time 14:15:30
set time hh:mm:ss
hh = hours
mm = minutes
ss = seconds
This command sets the internal clock (24-hour time). The example that
follows sets the internal time to 2:15 pm.
Note If seconds are omitted, the seconds default to 00.
Send:
set time 14:15
Receive:
set time 14:15 ok
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
addr dns
Receive:
addr dns 192.168.1.1
set addr dns address
This command sets the domain name server address, where address consists
of four numbers ranging from 0-255 inclusive, separated by ".".
Send:
set addr dns 192.168.1.1
Receive:
set addr dns 192.168.1.1 ok
C-Link Protocol Commands

Communications Configuration

Model 49i Instruction Manual
B-35

Advertisement

Table of Contents
loading

Table of Contents