C-Link Protocol Commands
Datalogging
B-18
Model 49i Instruction Manual
er xy
lr xy
sr xy
x = | 0 | 1 |
: Reply termination format (see "set format format"
command)
y = | 0 | 1 | 2 | : Output format (see "set erec/lrec/srec format format"
command)
These commands report the last long and short records stored or the
dynamic data record. In the example that follows, the command requests a
long record with no checksum, in ASCII format with text. For details on
how to decode the flag fields within these records, see
"flags" command.
Send:
lr01
Receive:
lr01
13:00 08-12-05 flags 1C00554A o3 0.000 hio3 0.000 cellai 0.000 cellbi
0.000 bncht 999.900 lmpt 999.900 o3lt 0.000 flowa 0.000 flowb 0.000
pres 0.000
erec
This command returns a brief description of the main operating conditions
at the time the command is issued (i.e. dynamic data). The example that
follows shows a typical response. The format is defined by the current
settings of "format" and "erec format" commands. For details on how to
decode the flag fields within these records, see
command.
Send:
erec
Receive:
erec
13:05 08-12-05 flags 1C00554A o3 0.000 1 lo o3 0.000 1 lampi 0000
bencht 999.900 lampt 999.900 ozlampt 0.000 flowa 0.000 flowb 0.000 Pres
0.000 avgt 10 lo avgt 10 O3bkg 0.000 O3 coef 1.000 lo O3 coef 1.000 O3
range 200000.000 lo O3 range 200000.000 int a 0000 int b 0000
sum 46f6
lrec
srec
lrec xxxx yy
srec xxxx yy
lrec aa:bb oo-pp-qq yy
srec aa:bb oo-pp-qq yy
xxxx = the number of past records
yy = the number of records to return (1 to 10)
aa = hours (01 to 24)
bb = minutes (01 to 59)
oo = month (01 to 12)
Figure B-1