C-Link Protocol Commands
Alarms
B-14
Model 49i Instruction Manual
set alarm level 1 value
set alarm level 2 value
set alarm level 3 value
set alarm level 4 value
set alarm level 5 value
These commands set the offset value (in ppb) from the concentration stored
at ozonator level 1-5 as the alarm trigger point for that level. The example
that follows sets the alarm level 1 offset to 20.5.
Send:
set alarm level 1 20.5
Receive:
set alarm level 1 20.5 ok
alarm oz lamp temp min
alarm oz lamp temp max
These commands report the ozonator lamp temperature minimum and
maximum value current settings. The example that follows reports that the
alarm ozonator lamp temperature minimum value is 52
Send:
alarm ozonator lamp temp min
Receive:
alarm ozonator lamp temp 52.0 deg C
set alarm oz lamp temp min value
set alarm oz lamp temp max value
These commands set the ozonator lamp temperature minimum and
maximum value to value, where value is a floating-point number
representing bench lamp temperature alarm limits in degrees C. The
example that follows sets the alarm ozonator lamp temperature maximum
value to 58
C.
°
Send:
set alarm ozonator lamp temp max 58.0
Receive:
set alarm ozonator lamp temp max 58.0 ok
alarm pressure min
alarm pressure max
These commands report the pressure alarm minimum and maximum value
current settings. The example that follows reports that the pressure alarm
minimum value is 200 mmHg.
Send:
alarm pressure min
Receive:
alarm pressure min 200 mmHg
C.
°