Thermo Scientific 49i Instruction Manual page 26

Uv photometric o3 analyzer
Table of Contents

Advertisement

Introduction
Specifications
1-4
Model 49i Instruction Manual
Power requirements
Physical dimensions
Weight
Analog outputs
Digital outputs
Digital inputs
Serial Ports
Ethernet connection
*
In non condensing environments. Performance specifications based on operation within 20–30 °C range.
100 VAC @ 50/60 Hz
115 VAC @ 50/60 Hz
220-240 VAC @ 50/60 Hz
150 watts
16.75" (W) X 8.62" (H) X 23" (D)
Approximately 35 lbs.
6 voltage outputs; 0
100 mV, 1, 5, 10 V (user selectable), 5% of
full-scale over/under range, 12 bit resolution, user selectable for
measurement input
1 power fail relay Form C, 10 digital relays Form A, user selectable
alarm output, relay logic, 100 mA @ 200 VDC
16 digital inputs, user select programmable, TTL level, pulled high
1 RS-232 or RS-485 with two connectors, baud rate 1200
data bits, parity, and stop bits, protocols: C-Link, MODBUS, and
streaming data (all user selectable)
RJ45 connector for 10Mbs Ethernet connection, static or dynamic
TCP/IP addressing
115200,
Thermo Fisher Scientific