run
up
1
2
3
4
button = | do | down | en | enter | he | help | le | left | me | menu | ri | right |
ru | run | up | 1 | 2 | 3 | 4 |
These commands simulates pressing the front panel pushbuttons. The
numbers represent the front-panel soft keys, from left to right.
Send:
push enter
Receive:
push enter ok
isc
iscreen
This command retrieves the framebuffer data used for the display on the
iSeries instrument. It is 19200 bytes in size, 2-bits per pixel, 4 pixels per
byte arranged as 320 by 240 characters. The data is sent in RLE encoded
form to save time in transmission. It is sent as a type '5' binary c_link
response with no checksum.
C-Link Protocol Commands