Record Layout
Definition
Thermo Fisher Scientific
9 flowa
10 flowb
11 pres
12 bncht
13 lmpt
relay
relay stat
This command reports the current relay logic normally "open" or normally
"closed," if all the relays are set to same state, that is all open or all closed.
The example that follows shows that the status when all the relays logic is set
to normally "open".
Send:
relay stat
Receive:
relay stat open
Note If individual relays have been assigned different logic, the response
would be a 4-digit hexadecimal string with the least significant byte (LSB)
being relay no 1.
For example:
Receive:
relay stat 0x0001 (indicates relay no 1 is set to normally
open logic, all others are normally closed)
Receive:
relay stat 0x0005 (indicates relay no 1 and 3 are set to be
normally open logic, all others are normally closed)
set relay open
set relay open value
set relay closed
set relay closed value
These commands set the relay logic to normally open or closed for relay
number value, where value is the relay between 1 and 16. The example that
follows sets the relay no 1 logic to normally open.
Note If the command is sent without an appended relay number then all the
relays are assigned the set logic of normally open/closed.
Send:
set relay open 1
Receive:
set relay open 1 ok
The Erec, Lrec, and Srec Layouts contain the following:
A format specifier for parsing ASCII responses
A format specifier for parsing binary responses
C-Link Protocol Commands

Record Layout Definition

Model 49i Instruction Manual
B-43