Servicing
Removing the Measurement Bench and Lowering the Partition Panel
Removing the
Measurement Bench
and Lowering the
Partition Panel
REMOVE SCREWS
Partition Panel
Loosen
Captive
Screw
Remove Screw
7-8
Model 42i Trace Level Instruction Manual
The measurement bench can be removed and the partition panel can be
lowered to improve access to connectors and components. Refer to the
following steps when a procedure requires lowering the partition panel (see
Figure
Pass Through
Connectors
Remove Screws
Figure 7–3. Removing the Measurement Bench and Lowering the Partition
Panel
Equipment Required:
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn instrument OFF and unplug the power cord.
7–3).
Remove Screw
LOWER PANEL
SLIDE BENCH OUT
Remove Hidden Screw