C-Link Protocol Commands
I/O Configuration
B-42
Model 49i Instruction Manual
Send:
dtoa 1
Receive:
dtoa 1 97.7%
Note All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
Table B-11. Default Output Assignment
D to A
Function
1
Voltage Output
2
Voltage Output
3
Voltage Output
4
Voltage Output
5
Voltage Output
6
Voltage Output
7
Current Output
8
Current Output
9
Current Output
10
Current Output
11
Current Output
12
Current Output
list var aout
list var dout
list var din
These commands report the list of index numbers and the variables
(associated with that index number) available for selection in the current
mode (determined by single/dual/auto) for analog output, digital output
and digital inputs. The index number is used to insert the variable in a field
location in a list using "set sp field index". The example that follows reports
the list of analog output, index numbers, and variables.
Send:
list var aout
Receive:
list var aout
index variable
0 none
1 o3
2 no2
5 cellai
6 cellbi
7 noisa
8 noisb