Thermo Scientific 49i Instruction Manual page 268

Uv photometric o3 analyzer
Table of Contents

Advertisement

C-Link Protocol Commands
Datalogging
B-20
Model 49i Instruction Manual
erec format
srec format
lrec format
These commands report the output format for long and short records, and
dynamic data in various formats such as ASCII without text, ASCII with
text, or binary. The example that follows shows the output format for long
records is ASCII with text, according to
Send:
lrec format
Receive:
lrec format 1
set erec format format
set lrec format format
set srec format format
These commands set the output format for long and short records, and
dynamic data, according to
long record output format to ASCII with text.
Send:
set lrec format 1
Receive:
set lrec format 1 ok
Table B-4. Record Output Formats
Format
Output Format
0
ASCII no text
1
ASCII with text
2
binary data
erec layout
lrec layout
srec layout
These commands reports the layout (string indicating the data formats) for
data that is sent out in response to the erec, lrec, srec, and related commands
The example that follows shows a typical response. For details on how to
interpret the strings, see "Record Layout Definition" later in this appendix.
Send:
lrec layout
Receive:
lrec layout %s %s %lx %f %f %f %f %f %f %f %f %f %f %f %f
t D L ffffffffffff
flags no nox hino hinox pres pmtt intt rctt convt smplf ozonf pmtv
Table
B-4.
Table
B-4. The example that follows sets the
Thermo Fisher Scientific

Advertisement

Table of Contents
loading