Thermo Scientific 49i Instruction Manual page 110

Uv photometric o3 analyzer
Table of Contents

Advertisement

Operation
Alarms Menu
Min and Max O
Temperature Limits
Lamp Temperature
Min and Max Lamp Temperature
3-74
Model 49i Instruction Manual
Lamp
The Minimum O
3
the minimum O
maximum O
In the Main Menu, choose Alarms > O
Use
Press
O3 LAMP TEMPERATURE:
ACTUAL MIN:
SET MIN TO:
RANGE
The Lamp Temperature screen displays the current lamp temperature and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 50 to 60 °C. If the lamp temperature reading goes beyond either
the minimum or maximum limit, an alarm is activated. The word
"ALARM" appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Lamp Temp.
LAMP TEMPERATURE:
ACTUAL
>MIN
MAX
RANGE
The Minimum Lamp Temperature alarm limit screen is used to change the
Limits
minimum lamp temperature alarm limit. The minimum and maximum
lamp temperature screens function the same way.
In the Main Menu, choose Alarms > Lamp Temp > Min or Max.
Lamp Temperature alarm limit screen is used to change
3
Lamp temperature alarm limit. The minimum and
3
Lamp temperature screens function the same way.
3
and
to increment or decrement the numeric value.
to save set to value as actual value.
60.0
65.0
INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
55.2
50.0
60.0
AVG
DIAGS ALARM
Lamp Temp > Min or Max.
3
o
C
o
C ?
o
C
o
C
o
C