LCD Module
Replacement
Thermo Fisher Scientific
2. Remove the three ribbon cables and the two-wire connector from the
front panel board.
3. Pop off the board from the two top mounting studs and remove the
board by lifting it up and off the slotted bottom support.
4. Replace the front panel board by following previous steps in reverse.
LCD Module Screws
Front Panel Board
Mounting Studs (2)
Ribbon Cables (3)
Figure 7-10. Replacing the Front Panel Board and the LCD Module
Use the following procedure to replace the LCD module
Equipment Required:
LCD module
Philips screwdriver
CAUTION If the LCD panel breaks, do not to let the liquid crystal contact
your skin or clothes. If the liquid crystal contacts your skin or clothes, wash
it off immediately using soap and water.
LCD Module
Bottom Support
Servicing

LCD Module Replacement

Two-Wire Connector
(Figure
7-10).