Operation
Alarms Menu

Intensity A and B

Min and Max Intensity Limits
3-78
Model 49i Instruction Manual
FLOW A:
ACTUAL MIN:
SET MIN TO:
RANGE
AVG
The Intensity A screen displays the current lamp intensity reading in Cell A
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 45,000 to 150,000 Hertz. If the intensity A reading goes beyond
either the minimum or maximum limit, an alarm is activated. The word
"ALARM" appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Intensity A or Intensity B.
INTENSITY A:
ACTUAL
>MIN
MAX
RANGE
AVG
The Minimum Intensity A alarm limit screen is used to change the
minimum intensity A alarm limit. The minimum and maximum intensity A
and intensity B screens function the same way.
In the Main Menu, choose Alarms > Select Intensity > Min or Max.
Use
and
Press
to save set to value as actual value.
0.400 LPM
0.500 LPM ?
INC/DEC
SAVE VALUE
DIAGS ALARM
98425 Hz
45000 Hz
150000 Hz
DIAGS ALARM
to increment or decrement the numeric value.