Operation
Alarms Menu
Intensity A and B
Min and Max Intensity Limits
3-78
Model 49i Instruction Manual
FLOW A:
ACTUAL MIN:
SET MIN TO:
RANGE
AVG
The Intensity A screen displays the current lamp intensity reading in Cell A
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 45,000 to 150,000 Hertz. If the intensity A reading goes beyond
either the minimum or maximum limit, an alarm is activated. The word
"ALARM" appears in the Run screen and in the Main Menu.
•
In the Main Menu, choose Alarms > Intensity A or Intensity B.
INTENSITY A:
ACTUAL
>MIN
MAX
RANGE
AVG
The Minimum Intensity A alarm limit screen is used to change the
minimum intensity A alarm limit. The minimum and maximum intensity A
and intensity B screens function the same way.
•
In the Main Menu, choose Alarms > Select Intensity > Min or Max.
•
Use
and
•
Press
to save set to value as actual value.
0.400 LPM
0.500 LPM ?
INC/DEC
SAVE VALUE
DIAGS ALARM
98425 Hz
45000 Hz
150000 Hz
DIAGS ALARM
to increment or decrement the numeric value.
Thermo Fisher Scientific