Thermo Scientific 49i Instruction Manual page 125

Uv photometric o3 analyzer
Table of Contents

Advertisement

Calibrate Pressure Span
Restore Default Pressure
Thermo Fisher Scientific
CALIBRATE PRESSURE ZERO:
CURRENTLY:
CONNECT VACUUM PUMP AND
RANGE
The Calibrate Pressure Span screen allows the user to view and set the
pressure sensor calibration span point.
Note The plumbing going to the pressure sensor should be disconnected so
the sensor is reading ambient pressure before performing the span
calibration. The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before calibrating.
In the Main Menu, choose Service > Pressure Calibration > Span.
Use
Press
CALIBRATE PRESSURE SPAN:
CURRENTLY:
RANGE
The Restore Default Pressure Calibration screen allows the user to reset the
Calibration
pressure calibration configuration values to factory defaults.
In the Main Menu, choose Service > Pressure Calibration > Set
Defaults.
Press
753.4 mmHg
SET TO:
0.0 mmHg ?
SAVE ZERO PRESSURE
AVG
DIAGS ALARM
,
,
and
to save set to value as actual value.
753.4 mmHg
SET TO:
760.0
MOVE CURSOR
CHANGE VALUE
AVG
DIAGS ALARM
to warn user and enable restore with
to move and change the value.
mmHg ?
SAVE
Model 49i Instruction Manual
Operation
Service Menu
.
3-89