Instrument Configuration - Thermo Scientific 49i Instruction Manual

Uv photometric o3 analyzer
Table of Contents

Advertisement

Set Analog Outputs

Instrument Configuration

Thermo Fisher Scientific
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-scale
voltage, zero sets the analog outputs to 0 volts, and normal operation. The
example below shows the selected output state "ALL" is set to normal.
In the Main Menu, choose Diagnostics > Test Analog Outputs > ALL,
Voltage Channel 1-6, or Current Channel 1-6.
Press
or
Press
to reset to normal.
SET ANALOG OUTPUTS:
SETTING:
OUTPUT SET TO:
SET TO FULL SCALE
SET TO ZERO
RESET TO NORMAL
RANGE
AVG
The Instrument Configuration screen displays information on the hardware
configuration of the instrument.
Note If the analyzer is in service mode, pressing
toggle it yes or no.
In the Main Menu, choose Diagnostics > Instrument Configuration.
Press
to toggle instrument configuration (in service mode only).
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD
SAMPLE/CAL VALVE
OZONATOR
DILUTION RATIO
AUTO CALIBRATION
RANGE
AVG
to set output.
ALL
NORMAL
DIAGS ALARM
YES
YES
YES
NO
NO
DIAGS ALARM
Operation
Diagnostics Menu
on the item will
Model 49i Instruction Manual
3-71