Thermo Scientific 49i Instruction Manual page 261

Uv photometric o3 analyzer
Table of Contents

Advertisement

Thermo Fisher Scientific
set alarm conc o3 min value
set alarm conc o3 max value
These commands set the O
value to value, where value is a floating-point representation of the
concentration alarm limits. The example that follows sets the O
concentration alarm maximum value to 150.
Send:
set alarm conc o3 max
Receive:
set alarm conc o3 max 150 ok
alarm intensity a min
alarm intensity a max
alarm intensity b min
alarm intensity b max
These commands report the cell a or b intensity alarm minimum and
maximum value current settings. The example that follows reports that the
cell a minimum value is 45000 Hz.
Send:
alarm intensity a min
Receive:
alarm intensity a min 45000 Hz
set alarm intensity a min value
set alarm intensity a max value
set alarm intensity b min value
set alarm intensity b max value
These commands set the cell a or b intensity alarm minimum and
maximum value to value, where value is a representation of the intensity
alarm limits. The example that follows sets the cell a intensity alarm
maximum value to 150000 Hz.
Send:
set alarm intensity a max 150000
Receive:
set alarm intensity a max 150000 ok
alarm level 1
alarm level 2
alarm level 3
alarm level 4
alarm level 5
These commands report the offset value (in ppb) from the concentration
stored at ozonator level 1-5 as the alarm trigger point for that level. The
example that follows reports that the alarm level 1 offset is 20.5.
Send:
alarm level 1
Receive:
alarm level 1 20.5
concentration alarm minimum and maximum
3
Model 49i Instruction Manual
C-Link Protocol Commands
Alarms
3
B-13