Setup Procedure

Thermo Fisher Scientific
7. Re-install the cover.
8. Remove any protective plastic material from the case exterior.
Use the following procedure to setup the instrument.
1. Connect the sample line to the SAMPLE bulkhead on the rear panel
(Figure
2-3). Ensure that the sample line is not contaminated by dirty,
wet, or incompatible materials. All tubing should be constructed of FEP
Teflon®, 316 stainless steel, borosilicate glass, or similar tubing with an
OD of 1/4-inch and a minimum ID of 1/8-inch. The length of the
tubing should be less than 10 feet.
Note Gas must be delivered to the instrument free of particulates. It may be
necessary to use the Teflon particulate filter as described in
Particulate Filter"
on
page
Note Gas must be delivered to the instrument at atmospheric pressure. It
may be necessary to use an atmospheric bypass plumbing arrangement as
shown in
Figure 2-4
if gas pressure is greater than atmospheric pressure.
2. Connect the EXHAUST bulkhead to a suitable vent. The exhaust line
should be 1/4-inch OD with a minimum ID of 1/8-inch. The length of
the exhaust line should be less than 10 feet. Verify that there is no
restriction in this line.
3. Connect a suitable recording device to the rear panel connector. See the
"Operation" chapter for more information about the rear panel
pin-outs.
4. Plug the instrument into an outlet of the appropriate voltage and
frequency.
WARNING The Model 49i is supplied with a three-wire grounding cord.
Under no circumstances should this grounding system be defeated.
9-2.
Model 49i Instruction Manual
Installation
Setup Procedure