Calculating / checking mAs values
Calculate the mAs values for 50Hz or 60 Hz according to
the following formulas:
•
For 50Hz:
Tube current (in A) x pulses x 0.02s = value (in mAs)
•
For 60Hz:
Tube current (in A) x pulses x 0.016s = value (in mAs)
The value obtained must be 28mAs ± 10%!
i
NOTE
If specified value cannot be obtained
Manual.
61 16 813 D3437
D3437.103.01.04.02
, see Service
8 Calculating / checking mAs values