Checking and Adjusting the X-Ray Beam
1. Place beam alignment tool vertically in the cassette carriage and move behind the slit cover plate.
2. Set the diaphragm 1 on the wheel.
To move the wheel, press button D.
X-radiation will only be activated when button D is engaged.
3. Turn unit ON and wait 1 min. for self-adjustment of the unit.
Thereafter select Service - Routine S.02 at the Multitimer. See next page.
14
.
2
X-RAY
P
I
8 0
kV
I 0
mA
Sirona Dental Systems GmbH
.
1
Beam
alignment
tool
59 01 603 D3285