9 Adjusting / calibrating the system

9.3 Checking the radiation field

1.
A
B
9.3
Checking the radiation field
i
NOTE
Illumination must be checked once the diaphragm has been
adjusted.
1. Clip distortion phantom A onto the X-ray detector cover.
2. Go to the
RADIATION FIELD
3. Make SIDEXIS XG ready for exposure:
Click
IMAGE ACQUISITION
The exposure dialog box showing the exposure status
appears in SIDEXIS XG.
Service routine S002.6 is displayed on the control pan-
el.
4. Press the R key to move the unit back to the starting
position.
5. Press the release button. Hold down the release button
while checking the distortion phantom.
Lighting strips B on the distortion phantom must not
light up!
26
sub-menu.
Maintenance Instructions GALILEOS
2.
2.
3.
4.
WAIT
UNTIL UNIT IS IN STARTING POSITION
5.
i
NOTE
If the strips on the phantoms light up at all it means overex-
posure of the system, and you cannot continue adjustment.
In this case, repeat the diaphragm adjustment procedure
and then check the radiation field again. If the lighting strips
still light up during the re-check of the radiation field, contact
the SIRONA Customer Service Center (KSC) to solve the
problem.
To confirm that the lighting strips on the distortion phantom
do not light up, click the check box underneath the message
window.
The box will appear checked.
Continue the calibration procedure with dosimetry.
i
NOTE
Leave the distortion phantom on the unit for the next calibra-
tion step.