3.10 Checking data paths / Generating test images

3 – 26
3.10

Checking data paths / Generating test images

SIDEXIS XG must be installed and configured before you begin checking the
data paths.
i
NOTE
For a panoramic test image, the panoramic mode must be activated on the
unit; for a ceph test image, the cephalometric mode must be activated; for a
TSA test image, the TS mode must be selected (see operating instructions).
Generate the test images for the different operating modes in succession.
1.
Select constancy test in SIDEXIS XG:
U
C
Select
TILITIES
ONSTANCY TEST
The classical SIDEXIS user interface is started. Constancy test is already
preset.
2.
Starting the exposure mode:
Click
XCXP
The dialog box for selecting the X-ray device opens.
i
NOTE
If no X-ray device has been configured yet in SIDEXIS XG, the password input
window appears instead of the dialog box for selecting the X-ray device.
3.
Select/confirm the X-ray device:
R
1
E.g. select
and click on
OOM
The dialog box for selecting the test type appears on the screen.
4.
Select/confirm the test type:
Click
SERVICE EXPOSURE
The dialog box for selecting the service exposure appears on the screen.
OK
59 38 399 D3352
D3352.076.01.26.02 09.2016
Tab 3