Konica Minolta cm-5 Instruction Manual page 96

Spectrophotometer
Hide thumbs Also See for cm-5:
Table of Contents

Advertisement

4. Use the
or
cursor to the desired item.
Settings
 OFF : Disable automatic white calibration (100%
Calibration) the next time the instrument is turned
ON (the power switch is set to ON).
 ON
: Enable automatic white calibration (100%
Calibration) the next time the instrument is turned
ON (the power switch is set to ON).
When you set automatic white calibration to ON, white
Memo
calibration (100% calibration) is automatically performed
after the power switch of the instrument is set to ON before
the <Welcome to Easy Setup Wizard!> screen appears.
If you have disabled the display of the <Welcome to Easy
Setup Wizard!> screen, the <Sample> screen is
displayed immediately after the instrument is turned ON
and white calibration (100% calibration) is performed. This
allows you to start measurement quickly. This function is
useful when you always perform measurement using the
same condition settings.
For the procedure of disabling the display of the <Welcome
to Easy Setup Wizard!> screen, refer to page 102.
5. Press the [OK/Edit] button.
The selection is confirmed and the screen returns to the
<Calibration Option Setting> screen.
If you press the [Back] button without pressing the [OK/
Notes
Edit] button, you return to the <Calibration Option
Settings> screen without changing the setting.
6. After you set the calibration options, press the [Back]
button to return to the <Configuration> screen.
E94
button of the cross key to move the