Specular Component (SCI/SCE)
When the measurement type is reflectance measurement, select the specular component mode.
When the Petri Dish measurement is selected, "SCE" is set automatically.
[Setting Procedure]
1. Use the
or
cursor to "SCI/SCE" and then press the [OK/Edit]
button.
The <Measurement Mode Settings> screen is displayed.
2. Use the
or
cursor to the desired item.
Settings
 SCI: Specular component included
 SCE: Specular component excluded
3. Press the [OK/Edit] button.
The selection is confirmed and the screen returns to the
<Measurement Condition Settings> screen.
If you press the [Back] button without pressing the [OK/
Notes
Edit] button, you return to the <Measurement Condition
Settings> screen without changing the setting.
E62
Start the procedure from the <Measurement Condition Settings> screen.
button of the cross key to move the
button of the cross key to move the