Connecting The Functional Ground Cable - Konica Minolta cm-5 Instruction Manual

Spectrophotometer
Hide thumbs Also See for cm-5:
Table of Contents

Advertisement

Receiving window of the Transmittance Specimen Chamber
1. Set measuring diameter to ø3 mm.
For the procedure for switching measuring diameter, refer to
Memo
page 61 "Measurement Area".
2. Use a blower to remove dirt and dust from the receiving
window.
Do not insert your finger through the receiving window to
Notes
touch the lens of the light-receiving optical system.

Connecting the Functional Ground Cable

Attach the Functional Ground Cable (CM-A211) in order to avoid communication problems with the PC caused by
Notes
static electricity.
[Procedure]
1. Connect the Y terminal of the Functional Ground Cable
to the Connection Terminal Screw on the rear of the
instrument.
2. Connect the other Y terminal of the Functional Ground Cable to a reliably grounded object (e.g.
a 100-240 VAC earth terminal or a metal desk).
Connection
Terminal Screw
Functional
Ground Cable
1
Receiving
window
Grounding
Terminal
Screw
E19