User Calibration - Konica Minolta cm-5 Instruction Manual

Spectrophotometer
Hide thumbs Also See for cm-5:
Table of Contents

Advertisement

User Calibration

You can perform calibration by using your own reference plate and calibration data instead of the white
calibration (100% calibration) data. The calibration data for user calibration can be specified by connecting the
instrument to a PC and using the optional Color Management Software "SpectraMagic™ NX".
You can select whether to use the user calibration data for measurement on the <Calibration Option Setting>
screen of the instrument.
[Procedure]
1. Press the [MENU] button.
The <Configuration> screen is displayed.
2. Use the
or
cursor to
calibration options and then press the
[OK/Edit] button.
The <Calibration Option Settings> screen is displayed.
3. Use the
or
cursor to the desired calibration data and then press the
[OK/Edit] button.
The <Calibration Options> screen is displayed.
4. Use the
or
cursor to "User Calibration" and then press the [OK/
Edit] button.
Now you can perform white calibration (100% calibration)
using your own calibration data.
You need to prepare user calibration data and write it to the
Notes
instrument memory before you can set user calibration to
ON.
E58
Start the procedure from the measurement screen.
button of the cross key to move the
button of the cross key to move the
button of the cross key to move the