<Transmittance measurement>
3. Open the Transmittance Specimen Chamber Cover and
set the Transmittance Zero Calibration Plate so that it
fully covers the illumination window.
Notes
Set the Transmittance Zero Calibration Plate horizontally,
as shown to the right. If the Transmittance Zero Calibration
Plate is set vertically, the Transmittance Specimen
Chamber Cover may not close.
4. Close the Transmittance Specimen Chamber Cover and
press the [OK/Edit] button.
0% calibration will be performed.
The Xe lamp flashes three times during the calibration.
During 0% calibration, the screen shows the number of flashes
performed.
When the 0% calibration is completed, the cursor moves to
100% Calibration.
<Liquid transmittance measurement>
3. Open the Transmittance Specimen Chamber Cover and
set the Transmittance Zero Calibration Plate so that it
fully covers the receiving window. (Refer to the figure
for the case of transmittance measurement.)
To use the 10 mm Wide Cell Measurement Set, set the
Transmittance Zero Calibration Plate (10 mm) so that it fully
covers the receiving window.
4. Close the Transmittance Specimen Chamber Cover and
press the [OK/Edit] button.
0% calibration will be performed.
The Xe lamp flashes three times during the calibration.
During 0% calibration, the screen shows the number of flashes
performed.
When the 0% calibration is completed, the cursor moves to
100% Calibration.
Notes
• After zero calibration (0% Calibration), you need to
perform white calibration (100% Calibration). Go to step
3 of "White Calibration (100% Calibration)" on page 55.
• Even though you performed white calibration (100%
Calibration) after turning ON the power, you need to
perform it again if you performed zero calibration (0%
Calibration) after the first white calibration.