Konica Minolta cm-5 Instruction Manual page 36

Spectrophotometer
Hide thumbs Also See for cm-5:
Table of Contents

Advertisement

8.
The <Target Mask Setting> screen
appears.
9. Attach the optional Target Mask
(for Petri Dish) if ø30mm was
selected or the optional Target Mask
(for Mini Petri Dish) if ø3mm was
selected in step 6 above.
10. Check that the cursor is on "Next",
and press the [OK/Edit] button.
11.
The <Meas. Condition Confirmation>
screen appears.
Check that the cursor is on "Next",
and press the [OK/Edit] button.
12.
The <Zero Calibration> screen
appears.
The following is the explanation when
zero calibration will be skipped.
13. Check that the cursor is on "Next",
and press the [OK/Edit] button.
14.
The <White Calibration> screen
appears.
Check that the cursor is on "White
Calibration", and press the [OK/
Edit] button.
White calibration is performed by using
the built-in White Calibration Plate of
the instrument.
E34
Memo
Target Mask
(for Petri Dish)
For the procedure of attaching
CM-A203
the Target Mask, refer to page
42.
Memo
Since the CM-5 stores the data
of the zero calibration
performed at the factory, it is
unnecessary to repeat the zero
calibration every time you
turn ON the instrument. For
details of zero calibration,
refer to page 52.
Memo
For details of white
calibration, refer to page 55.