11.
The <Specular Component Selection>
screen appears.
Use the
or
button of the cross
key to move the cursor to the desired
specular component and then press
the [OK/Edit] button.
The cursor moves to "Next".
12. Press the [OK/Edit] button.
13.
The <Meas. Condition Confirmation>
screen appears.
Check that the cursor is on "Next",
and press the [OK/Edit] button.
14.
The <Zero Calibration> screen
appears.
The following is the explanation when
zero calibration will be skipped.
15. Check that the cursor is on "Next",
and press the [OK/Edit] button.
16.
The <White Calibration> screen
appears.
Check that the cursor is on "White
Calibration", and press the [OK/
Edit] button.
White calibration is performed by using
the built-in White Calibration Plate of
the instrument.
Memo
Since the CM-5 stores the data
of the zero calibration
performed at the factory, it is
unnecessary to repeat the zero
calibration every time you
turn ON the instrument. For
details of zero calibration,
refer to page 52.
Memo
For details of white
calibration, refer to page 55.
2
E31