Error Messages

The following messages may appear while you are using the instrument. If such messages appear, take the
necessary actions shown in the table below. If the trouble does not go away in spite of taking the actions, contact
a KONICA MINOLTA-authorized service facility.
Notes
The table below shows messages that may be displayed on the LCD of the instrument. For communication error
check codes, refer to the separate document.
Message
Preparing to auto
select target color
ER005
Error in Flashing.
ER007
Clock Error
ER037
USER
CALIBRATION
DATA NOT SET
ER011
Calibration was not
performed correctly.
ER013
A/D Conversion
Error.
E160
Symptom/Possible Cause
(When Auto Target is set to ON)
• Instrument is preparing auto
target color settings.
Xenon lamp does not flash properly.
• Xenon lamp has blown.
• Faulty flash circuit.
• Faulty sensor
Clock IC is not working correctly.
• Since power voltage of backup
battery is exhausted due to
instrument not being used for a
long time, data of calendar or
clock has been lost.
• Internal backup battery has
reached the end of its life.
• Breakdown of circuits relating to
the clock IC.
Data for performing white
calibration or user calibration has
not been input in the instrument's
memory.
Zero calibration (0% calibration) or
white calibration (100% calibration)
has not been performed correctly.
Target Mask is not correct.
Failed during A/D conversion.
• Faulty A/D converter
• Breakdown of circuits relating to
the A/D converter
Corrective Action
Wait for a moment.
*This message will appear when the "Auto
Target" function is set to ON or when power
switch is turned ON with auto target set to ON.
When preparations are completed, this message
will disappear.
If this message keeps appearing, contact a
KONICA MINOLTA-authorized service facility.
Set date and time after charging the internal
backup battery. If this message keeps appearing,
the battery has reached the end of its life. Contact
a KONICA MINOLTA-authorized service
facility.
To input user calibration data to the instrument,
you need to connect the instrument with a
computer and use the optional software. For
details of the procedure, read the software
manual.
You cannot input the calibration data of the built-
in White Calibration Plate to the instrument. For
details, contact a KONICA MINOLTA-
authorized service facility.
Read page 52 "Zero calibration (0% Calibration)"
or page 55 "White Calibration (100%
Calibration)" and perform the calibration
properly such as by using the optional Zero
Calibration Box.
Attach the correct Target Mask.
Turn the power OFF, and then turn it ON again.
If this message keeps appearing, contact a
KONICA MINOLTA-authorized service facility.