Konica Minolta cm-5 Instruction Manual page 163

Spectrophotometer
Hide thumbs Also See for cm-5:
Table of Contents

Advertisement

Message
ER024
White calibration (100% calibration)
White Calibration
has not been completed.
has not been
performed
ER025
• Faulty drive motor
Lens Error
• Breakdown of drive circuit
• Faulty drive sensor
ER026
• Faulty drive motor
Shutter Error
• Breakdown of drive circuit
• Faulty drive sensor
ER027
Charging for flashing of xenon lamp
Charge Circuit is not
is not completed.
ready.
• Breakdown of charging circuit
ER031
• Faulty drive motor
Light Trap Error
• Breakdown of drive circuit
• Faulty drive sensor
ER036
User calibration has not been
User Calibration has
performed.
not been performed.
ER047
Measured reflectance or
Measurement error
transmittance values are abnormal.
ER069
Target color data which you
Data Protect is ON.
attempted to overwrite or delete is
protected.
ER083
The USB memory is not connected
USB memory stick
properly.
not connected
The USB memory cannot be
accessed due to security function.
The file was not saved in appropriate
file format.
ER091
Too many data are selected.
More than 31 data
are selected.
Symptom/Possible Cause
Corrective Action
The instrument requires white calibration (100%
calibration) for specified measurement conditions
when it is turned ON for the first time. (It is also
recommended that white calibration (100%
calibration) be performed whenever the
instrument is turned ON or when the ambient
environment changes.) White calibration (100%
calibration) may also be required when the
measurement area, specular component mode or
other measurement conditions are changed.
Confirm the measurement area, and then try the
measurement again. If this message keeps
appearing, contact a KONICA MINOLTA-
authorized service facility.
Confirm the shutter behavior, and then try the
measurement again. If this message keeps
appearing, contact a KONICA MINOLTA-
authorized service facility.
Confirm that
(Ready to measure) is displayed
and then perform measurement. If this message
keeps appearing, contact a KONICA MINOLTA-
authorized service facility.
Confirm the switching operation between SCI
and SCE, and then try the measurement again. If
this message keeps appearing, contact a
KONICA MINOLTA-authorized service facility.
When user calibration is to be performed instead of
white calibration (100% calibration), the instrument
requires user calibration for specified measurement
conditions when it is turned ON for the first time. (It
is also recommended that user calibration be
performed whenever the instrument is turned ON or
when the ambient environment changes.) User
calibration may also be required when the
measurement area, specular component mode or
other measurement conditions are changed.
Perform zero calibration and white/100% calibration
again according to the proper procedure, and then
re-measure. If message continues to appear, contact
a KONICA MINOLTA-authorized service facility.
If it is necessary to overwrite or delete the
protected target data, change the data protect
setting of the data to OFF.
Connect the USB memory properly and securely.
Disable the security function or use a USB
memory without security function.
A condition file must be created with the optional
software. For the procedure, refer to the software
manual.
Select only up to 31 data.
5
E161