Measurement

• Prior to the start of measurement, be sure to perform white calibration (100% calibration). For details, refer to
Notes
page 55 "White Calibration (100% Calibration)".
• To display the color difference, it is necessary to set target colors before measurement.
• For accurate measurement, make sure to keep measurement conditions (ambient temperature, etc.) constant.
[Operating Procedure]
1. When a screen other than the <Sample> screen is
displayed, press the [TARGET/SAMPLE] button.
The <Sample> screen is displayed.
• Every time the [Detail/List] button is pressed, the screen is
switched between the <Sample> detail screen and
<Sample> list screen.
Measurement can be performed on either the <Sample>
Memo
detail screen or <Sample> list screen.
2. Set the specimen on the Target Mask or Transmittance
Specimen Chamber according to the measurement type.
3. Make sure that
and then press the [MEAS] button.
The specimen is measured and the result is displayed on the
screen.
• The sample data No. is automatically assigned sequentially
in the order of measurements.
When the number of sample data sets stored in the memory
Memo
reaches 4000, the number assigned to the subsequent
sample data will always be No. 4000, and the oldest data
will be deleted one at a time.
E112
(Measurement possible) is displayed
Specimen