<Transmittance measurement>
4. Use the
or
button of the cross
key to move the cursor to the
"Transmittance" and then press the
[OK/Edit] button.
The cursor moves to "Next".
5. Press the [OK/Edit] button.
6.
The <Meas. Condition Confirmation>
screen appears.
Check that the cursor is on "Next",
and press the [OK/Edit] button.
7.
The <0% Calibration> screen appears.
The following is the explanation when
0% calibration will be skipped.
8. Check that the cursor is on "Next",
and press the [OK/Edit] button.
9.
The <100% Calibration> screen
appears.
Check that the Transmittance
Specimen Chamber is empty and
that the cursor is on "100%
Calibration", and press the [OK/
Edit] button.
100% calibration will be performed.
Memo
Since the CM-5 stores the data
of the 0% calibration
performed at the factory, it is
unnecessary to repeat the 0%
calibration every time you
turn ON the instrument. For
details of 0% calibration, refer
to page 52.
Memo
For details of 100%
calibration, refer to page
55.