<Reflectance measurement>
3. Press the [OK/Edit] button.
White calibration is performed by using the built-in White
Calibration Plate of the instrument.
The Xe lamp flashes three times during the calibration.
During white calibration, the screen shows the number of
flashes performed.
When the white calibration is completed, the screen returns to
the one displayed before the <Measurement Condition
Settings> screen.
<Reflectance measurement using Petri Dish (1)>
(Using the built-in White Calibration Plate to perform white calibration)
3. Press the [OK/Edit] button.
White calibration is performed by using the built-in White
Calibration Plate of the instrument.
The Xe lamp flashes three times during the calibration.
During white calibration, the screen shows the number of
flashes performed.
When the white calibration is completed, the screen returns to
the one displayed before the <Measurement Condition
Settings> screen.
<Reflectance measurement using Petri Dish (2)>
(Using the optional White Calibration Plate to perform white calibration as user calibration)
3. Insert the Calibration Glass for Petri Dish into the
recess in the Target Mask for Petri Dish.
4. Align the White Calibration Plate with the positioning
groove on the Target Mask for Petri Dish and push the
plate in.
5. Press the [OK/Edit] button.
White calibration will be performed.
The Xe lamp flashes three times during the calibration.
During white calibration, the screen shows the number of
flashes performed.
When the white calibration is completed, the screen returns to
the one displayed before the <Measurement Condition
Settings> screen.
E56
Calibration Glass
(for Petri Dish)
CM-A212
White Calibration
Plate
CM-A210