Auto Measurement

Measurement is repeated the specified number of times at the same position on the specimen and then the
average of the measured spectral reflectance/transmittance data is calculated. This will improve the accuracy of
the sample data.
Before starting auto averaging, you must complete the settings for auto averaging.
Notes
For details, refer to page 64 "Auto Measurement".
[Operating Procedure]
1. When a screen other than the <Sample> screen is
displayed, press the [Target/Sample] button.
The <Sample> screen is displayed.
• Every time the [Detail/List] button is pressed, the screen is
switched between the <Sample> detail screen and
<Sample> list screen.
Measurement can be performed on either the <Sample>
Memo
detail screen or <Sample> list screen.
2. Set the specimen on the Target Mask or Transmittance
Specimen Chamber according to the measurement type.
3. Make sure that
then press the [MEAS] button.
The number of completed measurements will be displayed
during measurement.
You cannot cancel the auto averaging in progress.
Notes
When the specified number of measurements has been
completed, the averaging result will be displayed on the
<Sample> screen as new sample data.
E136
(Ready to measure) is displayed and
Specimen