The <100% Calibration> screen
appears.
Place a specimen for 100%
calibration (a container containing a
liquid with high transmittance such
as distilled water) in the
Transmittance Specimen Chamber
10. Check that the cursor is on "100%
Calibration", and then press the
[OK/Edit] button.
100% calibration will be performed.
11.
When the 100% calibration is
completed:
If a USB memory device is not
connected, the <Meas. Cond. Settings
completed> screen appears. Skip to
step 12.
If a USB memory device is connected,
the < Save Condition File > screen
appears.
If the condition file will not be saved:
Check that the cursor is on "Next", and
press the [OK/Edit] button. The
<Meas. Cond. Settings completed>
screen appears. Skip to step 12.
To change the file name:
Use the
or
to move the cursor to the current file
name and press the [OK/Edit] button.
The <Edit Name> screen will appear.
Use the cross key buttons to move the
cursor to the desired character and press
the [OK/Edit] button to confirm it.
Press the [CAL] button to erase
characters.
When the file name has been input, press
the [MENU] button to move the cursor
to "Save", and press the [OK/Edit]
button to confirm the file name and
close the screen.
To change a file name using the USB
keyboard, remove the USB memory
device while on the name editing screen,
and input the name after connecting the
USB keyboard.
E40
button of the cross key
Cell or
Plastic Cell
Light
Memo
For details of 100%
calibration, refer to page
55.
Memo
A USB keyboard can also be
used to input file names. For
information on connecting a
USB keyboard, refer to page
143 "Connecting a USB
Keyboard".