Konica Minolta cm-5 Instruction Manual page 87

Spectrophotometer
Hide thumbs Also See for cm-5:
Table of Contents

Advertisement

Tolerance Setting
Specify the tolerance used for pass/fail judgment of measured data for each target color.
• Before tolerance is specified for individual target colors, the instrument is set with default tolerance. For details,
Memo
refer to page 89 "Default Tolerance Setting".
• The tolerance setting items are the same as those for the color space or index currently selected for the display
conditions.
• With the optional Color Management Software "SpectraMagic™ NX", you can set or use the color difference
tolerance easily.
[Setting Procedure]
1. Use the
or
button of the cross key to select the
"Current" tab, use the
cursor to "Tolerance", and then press the [OK/Edit]
button.
The <Tolerance> screen is displayed.
2. Use
and
or the
to move the cursor to the desired item.
3. Press the [OK/Edit] button to change the details.
• When the selected item is not currently checked, pressing
the [OK/Edit] button causes the item to be checked, and
you can change the setting value.
• When the selected item is not currently checked, pressing
the [OK/Edit] button causes the item to be checked. Press
button of the cross key to move the blue box cursor
the
to the setting value area. When you press the [OK/Edit]
button, the cursor appears on the value. Press the
button of the cross key to change the value.
You can move the cursor between the digits of the value by
Memo
pressing the
Settings
 0.1 to 20.0
* Settings of x and y are 0.001 to 0.200.
• You need to press the [OK/Edit] button to confirm the
change every time you change the setting of an item.
Start the procedure from the <Target Data Functions> screen.
or
button to move the
and
button of the cross key
or
button of the cross key.
or
2
E85