Shutter Open/Close
When cleaning the inside of the integrating sphere or to check the measurement points for a specimen using the
Specimen Viewing Mirror (Optional), you need to open or close the shutter of the specimen measuring port. Use
this option for the open/close operation.
During measurement, the shutter is automatically opened or closed according to the measurement type.
[Setting Procedure]
1. Use the
or
cursor to "Shutter Open/Close" and then press the
[OK/Edit] button.
Above and below the cursor  and  are displayed.
2. Use the
or
to open or close the shutter.
Settings
 Open: Opens the shutter
 Close: Closes the shutter
3. Press the [OK/Edit] button.
The shutter opens or closes according to the selection.
If you press the [Back] button without pressing the [OK/
Notes
Edit] button, you return to the <Measurement Options>
screen without changing the setting.
For the transmittance measurement, the geometry will be set to di:0° when the shutter is closed and will be set to
Memo
de:0° when the shutter is open and the Zero Calibration Box is set.
In the haze measurement, the optional Color Management Software SpectraMagic™ NX (CM-S100w) allows haze
calculation after automatic selection of the geometry (shutter open/close).
E66
Start the procedure from the <Measurement Options> screen.
button of the cross key to move the