<Sample> Detail Screen: "Difference" Tab
18 Sample data (The tab can be switched with the
Any color difference value which failed the pass/fail judgment based on the color difference
tolerance will be highlighted in red.
Pass/fail judgment: When the result was "Pass", the background is shown in green. When the result
was "Fail", the background is shown in yellow.
When the color assessment settings is set to ON, the tones are displayed. For the procedure of setting
the color assessment settings to ON/OFF, refer to page 97.
<Sample> Detail Screen: "Abs. & Diff." Tab
18 Sample data (The tab can be switched with the
• The upper section shows the sample data measured with illuminant 1, and the lower section shows
the sample data measured with illuminant 2. If illuminant 2 has not been set, "---" will be displayed.
Any color difference value which failed the pass/fail judgment based on the color difference
tolerance will be highlighted in red.
Pass/fail judgment: When the result was "Pass", the background is shown in green. When the result
was "Fail", the background is shown in yellow.
E114
or
button of the cross key.)
or
button of the cross key.)
18
18