Troubleshooting - Konica Minolta cm-5 Instruction Manual

Spectrophotometer
Hide thumbs Also See for cm-5:
Table of Contents

Advertisement

Troubleshooting

If an abnormality has occurred with the instrument, take necessary actions as given in the table below. If the
instrument still does not work properly, turn the power OFF, and then turn it ON again. If the symptom remains,
contact a KONICA MINOLTA-authorized service facility.
Symptom
LCD is blank.
Is the AC adapter connected?
Measuring button is
Is measurement still in progress?
not effective.
Is a screen where measurement is
possible displayed?
Abnormal
Is the specimen placed properly?
measurement result
Is proper calibration data used?
Was white calibration (100%
calibration) performed correctly?
Was zero calibration (0%
calibration) performed correctly?
Measurement results
Is the instrument and specimen kept
fluctuate.
stationary during measurement?
Not possible to input
Is the USB cable connected
data to computer.
correctly?
No commands from
Is the USB cable supplied with the
computer are
instrument used?
accepted.
Commands cannot
be accepted
correctly.
Files cannot be saved
Is the USB memory connected
from the instrument
properly?
to a USB memory.
Is
Files cannot be read
shown on the LCD?
from a USB memory
to the instrument.
Not possible to print. Is the printer cable connected
properly?
Check Point
(USB memory connected)
Action
Connect the AC adapter.
Wait until measurement is complete, and then
press the button.
Measuring button must be pressed while a screen
that allows measurement (e.g. Calibration,
Target, or Sample screen) is shown.
Make sure that the specimen is securely placed
in the instrument to prevent leakage of light.
To perform user calibration or Petri Dish
measurement, you need to input appropriate
calibration data to the instrument by using the
optional software. Check that the calibration
plate is correct for the input calibration data
before performing white calibration (100%
calibration) or user calibration.
Read page 52 "Zero calibration (0% Calibration)"
and perform the calibration properly such as by
using the optional Zero Calibration Box.
Do not allow the instrument and specimen to
move during measurement.
Connect the instrument's USB connecting
terminal to the computer's USB port with the
USB cable supplied with the instrument.
Connect the USB memory to the USB connection
terminal of the instrument properly and securely.
If the problem continues, turn OFF the
instrument once and turn it ON again, and then
connect the USB memory.
Connect the RS-232C connector of the
instrument and the connector of the printer with
the optional connecting cable or the cable
conforming to the requirements shown in the
connection diagram on page 146.
5
E163

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents