Displayed Average Noise Level (Danl) - Keysight Technologies X Series Manual

Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
2

Displayed Average Noise Level (DANL)

Test Limits (with 0 dB input attenuation)
See
The Displayed Average Noise Level (DANL) of the signal analyzer is measured
across a 10 kHz frequency span at several center frequencies. The analyzer
input is terminated into a 50Ω load. A test is performed to assure the
measurement is not performed in the presence of a residual response. The
measurement is then averaged, and the result is normalized to a 1 Hz
bandwidth.
Item
Termination, 50Ω Type-N(m)
Termination, 50Ω 2.4 mm (f)
a. for Options 503, 508, 513, and 526
b. for Options 543, 544, and 550
Figure 2-1
DANL Test Setup
Table 2-1 on page 13
Critical Specifications
(for this test)
Frequency: 10 MHz to 18 GHz
a
Frequency: 10 MHz to 50 GHz
b
for values.
11
Recommended
Keysight Model
909A Option 012
85138B