Function Description
7.1

Manual Ventilation

Figure 16 Manual Ventilation (Inspiration) - Typical for Breathing System P/N 4116398 or 4117529
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Manual Ventilation: General
During manual ventilation, the APL valve is set to the "MAN" position. The
patient system safety valve is activated. The piston of the ventilator is in the
upper end position in order to reduce the volume of the ventilator.
Manual Ventilation: Inspiration
During inspiration, expiratory valve remains closed. When the clinician com-
presses the breathing bag the gas mixture (expiratory gas and fresh gas)
flows through the fresh-gas decoupling valve, the inspiratory valve, the O2
sensor, the inspiratory hose, and the Y-piece into the patient's lung. The pres-
sure sensor measures the airway pressure. The APL valve limits the ventila-
tion pressure. Any excess amount of the gas mixture flows through the APL
valve and the non-return valve to the anesthetic gas scavenging system.
6020.002
Fabius Tiro
Revision B Released
25