Figure 20 PEEP Control Assembly
184
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
16. If Software version 2.N is installed, perform the
dure in accordance with the Adjustment and Calibration Section.
17. Perform the PMS procedure.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Replacement Procedure
Pump Calibration
Revision B Released
proce-