Ventilation - Dräger Medical Oxylog 3000 Technical Documentation Manual

Emergency and transport ventilator
Table of Contents

Advertisement

The "!!!Supply pressure low" message is no longer displayed.
3.9

Ventilation

(√ )
3.9.1
Volume-controlled ventilation
1. Connect unit to power and to O2 compressed gas supply.
2. Connect ventilation tube and ventilation valve to flow sensor and
flowmeter tubes.
3. Connect test specimens (breathing bag and catheter socket) to flow
sensor with elbow adapter.
4. Make the following settings:
5. Press the "Values" button to select the measured values MV and VTE.
The displayed MV is in the range 1.3 L/min to 2.3 L/min.
6. Set VT to 500 mL.
The displayed MV is in the range 3,7 L/min to 5,3 L/min.
7. Set VT to 1000 mL.
The displayed MV is in the range 8 L/min to 10 L/min.
(√ )
3.9.2
Pressure-controlled ventilation
1. Remove test specimens and connect test thorax in their place.
2. Make the following settings:
5503.403
Oxylog 3000
Ventilation mode = IPPV
Frequency = 10 1/min
Alarm limit Pmax = 60 mbar
PEEP = 5 mbar
O2 = 60 Vol.%
I/E = 1:1
TPlat = 50%
Trigger = off
VT = 200 mL
Ventilation mode = BIPAP
Frequency = 6 1/min
Alarm limit Pmax = 60 mbar
Released
2002-03-07
[OK]
[OK]
[OK]
[OK]
7/11