O2 Flush Valve - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Replacement Procedure
7

O2 Flush Valve

Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
The O2 flush valve is accessible from the underside of the table top. Valve
removal requires disconnecting the plastic tubing and removal of the fittings.
Figure 7
shows mounting and tubing arrangement for the valve.
1. Invert unit by performing steps given in the
sembly
Section.
2. Disconnect the tubing from the fittings on the O2 flush valve.
3. Remove the fittings from the outlet of the O2 flush valve.
4. Un-screw the plastic retaining nut from the valve body, and withdraw the
valve from the front of the unit.
5. Install the replacement valve in the unit, and secure it with the plastic
retaining nut.
6. Install fittings from the removed O2 flush valve onto the replacement O2
flush valve.
7. Reconnect the tubing to the fittings as shown in
8. Install the appropriate label on the replacement valve.
9. Invert unit into its upright position by performing steps given in the
Module Inversion Reassembly
10. Perform the PMS Procedure.
6020.002
Core Module Inversion Disas-
Figure
7.
Section.
Revision B Released
Fabius Tiro
Core
155