Dräger Medical Fabius Tiro Technical & Service Manual page 41

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Function Description
Figure 25 Volume Control Ventilation (Inspiration) - Typical for Breathing System P/N 4118378 or 4118379
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Volume/Pressure Control Mode: Expiration
During expiration, the inspiratory valve remains closed thus preventing
rebreathing into the inspiratory branch.
The expiratory gas from the lung flows through the expiratory hose, the flow
sensor, the PEEP/Pmax valve, the expiratory valve, and through the absorber
back into the breathing bag mixing with fresh gas also flowing into the breath-
ing bag.
The ventilator's piston moves back drawing the gas mixture needed for the
next inspiration into the piston space.
6020.002
Fabius Tiro
Revision B Released
35