Spontaneous Breathing - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Fabius Tiro
Figure 19 Manual Ventilation (Expiration) - Typical for Breathing System P/N 4118378 or 4118379
7.2
Spontaneous
Breathing
28
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Spontaneous Breathing: General
A prerequisite for spontaneous breathing is that the patient is supplied with a
sufficient amount of fresh gas. The APL valve selector must be set to the
"SPONT" position. No pressure builds up in the compact breathing system.
Spontaneous Breathing: Inspiration
During inspiration, the expiratory valve remains closed thus preventing
rebreathing of expiratory gas containing CO2.
6020.002
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description
Revision B Released