Pressure Calibration - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Fabius Tiro
6

Pressure Calibration

Zero Calibration
216
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Adjustment and Calibration Procedures
Figure 6
Oxygen Monitor Service Screen
1. Set the main power switch to ON.
2. Access the main service screen (ref.
3. Scroll to Calibration and select Pressure to bring up the pressure calibra-
tion screen. See
Figure
a) Remove the pressure sample line and ventilator hose from the
breathing system, and expose the sample line to air.
b) Let the current pressure value stabilize.
c) Touch the 'Store Zero' softkey to enter the current value as the cali-
bration zero.
4. Re-connect the pressure sample line and the ventilator hose to the
breathing system.
5. Press the Standby key to return to the standby screen.
6002.002
RETURN TO CD-ROM TABLE OF CONTENTS
Main Service Screen
7.
Revision B Released
Section).