PMS PROCEDURE (continued)
( )
8.1.4.3
( )
8.2

System Diagnostics

8.2.1
Connect the pipeline supply or open the cylinders.
8.2.2
Push the System Power switch to ON ( ).
8.2.3
Verify that the following is displayed and system diagnostics indicate a
"Pass" condition:
SYSTEM DIAGNOSTICS
SYSTEM DIAGNOSTICS
Watch Dog
Watch Dog
System RAM
System RAM
Program Memory
Program Memory
Video Test
Video Test
Interrupts
Interrupts
A/D Converter
A/D Converter
NV RAM
NV RAM
Serial Port
Serial Port
Clock
Clock
Speaker
Speaker
Main Power
Main Power
Battery
Battery
( )
8.2.4
Record the machine software version on the header of the checklist form.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
APPENDIX A
Record the total leakage current with the Polarity and Ground
switches set as follows:
Ground
Polarity
Open
Normal
Normal
Normal
Open
Reversed
Normal
Reversed
Verify that the leakage current is 300 microamps or less in each of
the switch positions (500 microamps or less for the 220/240 volt
power supply option).
Fabius Tiro SW XX
Fabius Tiro SW 1.03 CRC 6DBA
18
RETURN TO CD-ROM TABLE OF CONTENTS
System Status
System Status
Pass
Pass
Functional
Functional
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
FABIUS TIRO