Fabius Tiro Parts Replacement Schedule - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

FABIUS TIRO

Fabius Tiro Parts Replacement Schedule:

Quantity
1
Flow Sensor (5 pack)
1
O2 Sensor
2
O-ring, Vent Hose
2
Valve Disc
1
Hose asm, PEEP/Pmax - APL Bypass
1
Hose, Pressure
1
Filter, AGS
Annually: Fabius GS/Tiro 1 year kit
2
Filter, Pressure & Pneumatic asm
1
Diaphragm, Patient
4
O-ring, Vapor
1
O-ring, for Diaphragm (piston)
Every 3 Years: Fabius GS/Tiro 3 year kit
1
Diaphragm, Piston
2
Battery, 12V Rechargeable
1
Canister Assembly
1
Lip Seal
1
Packing Ring
1
PM Kit; 1 Year
Disposal of used batteries and O2 Sensors:
• Batteries must be disposed of in conformity with local waste disposal regulations.
• Expired O2 sensors can be returned to:
Dräger Medical AG & Co. KGaA
Moislinger Allee 53-55
Reparaturannahme
23542 Lübeck
Germany
• Do not open forcibly: danger of chemical burns.
• Do not incinerate: danger of explosion.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
APPENDIX A
Description
As Required
3
PMS PROCEDURE (continued)
Alternate
Part Number
Part Number
8403735
6850645
2M08777
2123249 or
M23225
4117027
1190520
M33294
4117360-001
4199912
8402868
2600650
4115864
8604831
4117360-002
4199911
2600651
4114229
M29320
M30455
M30456
4117360-001
4199912