Vacuum Pressure Pump Assembly (Software Version 2.N); Pump Calibration - If Applicable - Dräger Medical Fabius Tiro Technical & Service Manual

Anesthesia system
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Adjustment and Calibration Procedures
10
Vacuum Pressure
Pump Assembly
(Software Version
2.N)
10.1
Pump Calibration - If
Applicable
Dräger Medical AG & Co. KGaA
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NOTE
The ability to calibrate the pump is dependant upon the current PCB
installed in the Core Unit.
1. Access the 'System Service Screen' screen and note the installed 'Hard-
ware Part Number'. Refer to
2. If the 'Hardware Part Number' listed is 4118079, proceed with the next
step. If the 'Hardware Part Number' is 4116632, the pump calibration pro-
cedure is not available.
3. Access the 'Real Time Values' screen, Refer to
tion.
NOTE
Upon initial entry into this screen, the calibration operation is not active.
NOTE
The pump calibration operation can be performed with the pump in either
the ON or OFF state. However, after the calibration operation is complete,
the pump will remain in the ON state unless manually turned off. Refer to
Pump On/Off
Section.
4. Press the 'PUMP CAL' softkey to activate the pump calibration operation.
The 'PUMP CAL' label will be highlighted in reverse upon activation and
will remain in this condition until the calibration operation is either com-
plete or aborted.
6002.002
System Service Screen
Section.
Real Time Values
Revision B Released
Fabius Tiro
Sec-
221